Pottstown Skilled Nursing And Rehabilitation Cente
Pottstown Skilled Nursing and Rehabilitation Cente in POTTSTOWN, PA — inspection on February 24, 2026.
Found 1 citation. Severity: Standard violations.
Health inspections identify deficiencies that facilities must correct within required timeframes. Violations range from minor documentation issues to serious safety concerns and are subject to follow-up verification.
Inspection Findings
record review revealed that Resident 1 had diagnoses that included congestive heart failure, presence
physician's orders dated January 23 through 29, 2026, and the Treatment Administration Record for January 2026, revealed the following: Staff were to perform a self-test to ensure proper function of the left ventricular assist device (LVAD) every shift.
There was no documented evidence that staff performed the self-test on two occasions.
Staff were to record pump rate, pulse index, pump power, pump speed, and vital signs.
Staff were to use a doppler to obtain mean blood pressure two times a day.
There was no documented evidence that the rate, index, power, speed, vital signs, or mean blood pressure were obtained on six occasions.
Staff were to ensure backup batteries were on charge at all times and verify placement every shift.
There was no documented evidence that staff performed this task on two occasions.
Staff were to monitor daily weights and notify the provider of changes greater than two pounds (lbs.) in one day or five lbs. in one week.
There was no documented evidence that staff obtained the resident's weights on four occasions.
Staff were to convert to batteries one time a day and at bedtime return to main power.
There was no documented evidence that staff converted to batteries two occasions.
Staff were to observe for signs and symptoms of driveline infection every shift.
There was no documented evidence that staff observed for signs and symptoms of driveline infection on three occasions.
Staff were to ensure that if the resident goes out of the building, the backup controller, two extra batteries, and two extra clips go out with the resident. On January 26, 2026, the resident was transferred to the hospital for a change in condition.
There was no documented evidence that the resident was sent with the backup controller, two extra batteries, and two extra clips. In an interview on February 24, 2026, at 3:00 p.m., the Director of Nursing confirmed that there was no evidence that staff implemented the physician's orders as identified. 28 Pa.
Code 211.12(d)(1)(5) Nursing services.
Any deficiency statement ending with an asterisk (*) denotes a deficiency which the institution may be excused from correcting providing it is determined that other safeguards provide sufficient protection to the patients. (See instructions.) Except for nursing homes, the findings stated above are disclosable 90 days following the date of survey whether or not a plan of correction is provided.